B Class Autoclaves are advanced steam sterilizers used in medical, dental, and laboratory settings. They are the highest standard of tabletop autoclaves and meet European EN 13060 standards for sterilization.

Features:
-
- SS 316 medical-grade steel chamber
- Aluminium instrument tray
- European Class B Standard with 4-step vacuum drying
- Inbuilt steam generator
- ARM-based microcontroller system
- Safety: Overheat control, overpressure valve, door lock system
- Thermal printer for data logging
Technical Specifications:
Parameter |
16 Litres |
|---|---|
| Product Dimensions (W×H×D) | 500mm × 440mm × 680mm |
| Chamber Diameter | 230mm |
| Chamber Length | 370mm |
| Base Bush (W×H) | 330mm × 490mm |
| Max Load (Vacuum) | 4.5 kg |
| Max Load (Non-Vacuum) | 2.5 kg |
| Water Tank Capacity | 5 Litres |
| Instrument Rack Size (L×W×H) | 11.5″ × 7.0″ × 6.5″ |
| Operating Temperature | 121°C – 134°C |
| Operating Pressure | 0.2 – 2.45 Bar |
| Voltage | 230V / 50Hz |
| Noise Level @ 1m | < 60 dB |
| Chamber & Lid Material | SS 316 |
| Tray Material | Aluminium 6061 |
Class B Sterilization Cycle (with Pressure & Vacuum Phases)
The Class B Autoclave follows a three-phase sterilization cycle that ensures complete penetration of steam, even inside hollow and wrapped instruments.
1. Pre-Vacuum Phase (0 to -0.9 bar)
Multiple vacuum pulses are created to remove air from the chamber. This drops the pressure to approximately -0.9 bar, allowing saturated steam to fully reach internal surfaces of instruments.
2. Sterilization Phase (+2.1 to +2.45 bar)
Steam is introduced and maintained at a constant pressure between 2.1 and 2.45 bar, depending on the selected cycle (121°C or 134°C). This kills all microorganisms, including spores.
3. Post-Vacuum & Drying Phase (-0.8 to -0.9 bar)
A final vacuum removes moisture from the load. Pressure drops again to around -0.9 bar, ensuring dry, sterile instruments ready for immediate storage or use.
Pressure Summary:
Phase |
Approx. Pressure |
|---|---|
| Pre-Vacuum | -0.8 to -0.9 bar |
| Sterilization | +2.1 to +2.45 bar |
| Drying (Post-Vac) | -0.8 to -0.9 bar |
Pre-Programmed Sterilization Modes (Both Models)
Mode |
Sterilization Time |
Dry Time |
Total Time |
|---|---|---|---|
| 121 Fast | 18 min | 3 min | 28 min |
| 121 Wrapped Solids | 18 min | 5 min | 30 min |
| 134 Fast | 5 min | 3 min | 20 min |
| 134 Wrapped Solids | 5 min | 5 min | 23 min |
| 134 Hollow Open | 5 min | 5 min | 23 min |
| PRION (134°C) | 20 min | 5 min | 40 min |
| Helix/Bowie-Dick Test | 3.5 min | 4 min | 28 min |
| Vacuum Test | – | – | 15 min |
What’s Different in 16 & 23 litres?
Only the physical dimensions, chamber size, and load capacity vary between the two:
Model |
Chamber Volume |
Max Load |
|---|---|---|
| 16 Litre | 16 Litres | 4.5 kg (vacuum mode) |
| 23 Litre | 23 Litres | 6.0 kg (vacuum mode) |